1、Keysight TechnologiesE4416A/E4417A EPM-P SeriesPower Meters and E-Series E9320Peak and Average Power SensorsData Sheet02 | Keysight | E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors Data SheetEPM-P Power Meter SpecificationsSpecifications describe the instru
2、ments warranted performance and apply after a 30 minute warm-up. These specifications are valid over its operating and environmental range unless otherwise stated and after performing a zero and calibration procedure.Supplemental characteristics are intended to provide additional information; useful
3、 in applying the instrument by giving typical (expected), but not warranted performance parameters. These characteristics are shown in italics or labeled as typical, nominal or approximate.Measurement uncertainties information can be found in, Fundamentals of RF and Microwave Power Measurements, App
4、lication Note 64-1, literature number 5965-6630E.Compatibility, the EPM-P series power meters operate with the E-series E9320 family of power sensors for peak, average and time-gated power measurements. The EPM-P series also operates with the existing 8480 and N8480 series, E-series CW and the E9300
5、 range of power sensors for average power measurements. For specifications pertaining to the 8480 and E-series CW and E9300 power sensors, please refer to the EPM Series Power Meters, E-Series and 8480 Series Power Sensors, Technical Specifications, literature number 5965-6382E. For specifications p
6、ertaining to the N8480 series power sensors, please refer to the N8480 Series Thermocouple Power Sensors, Technical Specifications, literature number 5989-9333EN.Measurement modes, the EPM-P series power meters have two measurement modes: Normal mode (default mode using E9320 sensors) for peak, aver
7、age and time-related measurements, and Average only mode. This mode is primarily for average power measurements on low-level signals, when using E9320 sensors, and is the mode used with 8480 and N8480 series sensors, E-series CW sensors and E-series E9300 sensors.Frequency range: 9 kHz to 110 GHz, s
8、ensor dependentPower range: 70 to +44 dBm, sensor dependent03 | Keysight | E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors Data SheetSingle Sensor Dynamic RangeE-series E9320 peak and average power sensors 70 dB maximum (normal mode);85 dB maximum (average o
9、nly mode)E-series CW power sensors 90 dBE-series E9300 average power sensors 80 dB maximum8480 series sensors 50 dB maximumN8480 series sensors 55 dB maximumDisplay unitsAbsolute Watts or dBmRelative Percent or dBDisplay resolution Selectable resolution of 1.0, 0.1, 0.01, 0.001 dB in logarithmic mod
10、e, or 1 to 4 significant digits in linear mode.Offset range 100 dB in 0.001 dB increments, to compensate for external loss or gainVideo bandwidth 5 MHz (set by meter and is sensor dependent)Note that the video bandwidth represents the ability of the power sensor and meter to follow the power envelop
11、e of the input signal. The power envelope of the input signal is, in some cases, determined by the signals modulation bandwidth, and hence video bandwidth is sometimes referred to as modulation bandwidth. The PCIe x16 Gen 3 link between CPU and backplane switch provides up to 16 GB/s theoretical, 14
12、.2 GB/s practical.Video bandwidth/dynamic range optimizationThe power measurement system, comprising the sensor and meter, has its maximum video bandwidth defined by the E9320 sensor. To optimize the systems dynamic range for peak power measurements, the video bandwidth in the meter can be set to Hi
13、gh, Medium and Low, as detailed in the following table. The filter video bandwidths stated in the table are not the 3 dB bandwidths as the video bandwidths are corrected for optimal flatness. Refer to figures 6 to 8 for information on the sensors peak flatness response. A filter OFF mode is also pro
14、vided.Table 1. Video bandwidth versus peak power dynamic range Sensor model Video bandwidth/maximum peak power dynamic ranOFF High Medium LowE9321A E9325A300 kHz/ 40 dBm to +20 dBm 300 kHz/ 42 dBm to +20 dBm100 kHz/ 43 dBm to +20 dBm30 kHz/ 45 dBm to +20 dBmE9322A 36 dBm to +20 dBm1.5 MHz/ 37 dBm to
15、 +20 dBm1.5 MHz/ 38 dBm to +20 dBm300 kHz/ 39 dBm to +20 dBm100 kHz/E9326A 39 dBm to +20 dBmE9323A E9327A5 MHz/ 32 dBm to +20 dBm5 MHz/ 32 dBm to +20 dBm1.5 MHz/ 34 dBm to +20 dBm300 kHz/ 36 dBm to +20 dBm04 | Keysight | E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Pow
16、er Sensors Data SheetAccuracyInstrumentationPlease add the corresponding power sensor linearity percentage; see Tables 6a and 6b for the E9320 sensors.Average only mode Absolute Logarithmic 0.02 dBLinear 0.5%Relative Logarithmic 0.04 dBLinear 1.0%Normal mode Calibration temperature1 5 C Temperature
17、0 to 55 CAbsolute accuracy (log) 0.04 dB 0.08 dBAbsolute accuracy (linear) 0.8% 1.7%Relative accuracy (log) 0.08 dB 0.16 dB Relative accuracy (linear) 1.6% 3.4%Time base accuracy 0.01%1 mW power referencePower output 1.00 mW (0.0 dBm). Factory set to 0.5% traceable to the National Physical Laborator
18、ies (NPL), UK2Accuracy For two years 0.5% (23 3 C) 0.6% (25 10 C) 0.9% (0 to 55 C)Frequency 50 MHz nominalSWR 1.06 maximum (1.08 maximum for Option E41xA-003)Connector type Type N (f), 50 ohmsMeasurement CharacteristicsMeasurements Average powerPeak powerPeak-to-average ratioMeasurements between two
19、 time offsets (time-gating)Averaging Averaging over 1 to 1024 readings is available for reducing noiseMeasurement speed (GPIB) Over the GPIB, three measurement speeds are available (normal, x 2 and fast). The typical maximum speed is shown in the following table.1. Power meter is within 5 C of its c
20、alibration temperature.2. National metrology institutes of member states of the Metre Convention, such as the National Institute of Standards and Technology in the USA, are signatories to the Comit International des Poids et Mesures Mutual Recognition Arrangement. Further information is available fr
21、om the Bureau International des Poids et Mesures, at http:/www.bipm.fr/ 05 | Keysight | E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors Data SheetTable 2. Measurement speed for different sensor typesSensor type Measurement speed (readings/second)Normal x 2 F
22、ast1,2E-Series E9320 peak and average sensorsAverage only mode 20 40 400Normal mode320 40 1000E-Series CW and E9300 average power sensors20 40 4008480 and N8480 Series sensor 20 40 N.A.Channel functions A, B, A/B, B/A, A-B, B-A and RelativeStorage registers 10 instrument states can be saved via the
23、Save/Recall menu.Predefined setups For common wireless standards(GSM900, EDGE, NADC, iDEN, Bluetoothnull, IS-95 CDMA, W-CDMA and cdma2000null), predefined setups are provided.TriggerSources Internal, External TTL, GPIB, RS232/422Time resolution 50 nsDelay range 1.0 sDelay resolution 50 ns for delays
24、 2.0 V, Low 1 kWTrigger out Output provides TTL compatible levels (high 2.4 V, low 0.4 V) and uses a BNC connector1. Fast speed is not available for 8480 and N8480 series sensors.2. Maximum measurement speed is obtained by using binary output in free run trigger.3. For E9320 sensors, maximum speed i
25、s achieved using binary output in free run acquisition.06 | Keysight | E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors Data SheetSampling CharacteristicsSampling rate 20 Msamples/secondSampling technique Continuous samplingRear Panel Inputs/OutputsRecorder o
26、utput(s) Analog 0 to 1 V, 1 kW output impedance, BNC connector. Two outputs are available on E4417A (channels A and B).Remote input/outputTTL output Used to signal when measurement has exceeded a defined limitTTL input Initiates zero and calibration cycleConnector type RJ-45 series shielded modular
27、jack assemblyTTL output High = 4.8 V max; Low = 0.2 V max.TTL input High = 3.5 V min, 5 V max; Low = 1 V max, 0.3 V minRS-232/422 interface Serial interface for communication with an external controller. Male plug 9-pin D-subminiature connector.Trigger in Accepts a TTL signal for initiating measurem
28、ents, BNC connectorTrigger out Outputs a TTL signal for synchronizing with external equipment, BNC connectorGround Binding post accepts 4 mm plug or bare wire connectionLine power Input voltage range 85 to 264 Vac, automatic selectionInput frequency range 47 to 440 HzPower requirement Approximately
29、50 VA (14 Watts)Remote ProgrammingInterface GPIB interface operates to IEEE 488.2 and IEC-625. RS-232 and RS-422 serial interfaces supplied as standardCommand language SCPI standard interface commandsGPIB compatibility SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP1, DC1, DT1, C0.Environmental Specificati
30、onsOperating environmentTemperature 0 to 55 CMaximum humidity 95% at 40 C, (non-condensing) Maximum altitude 3,000 meters (9,840 feet)Storage conditionsStorage temperature 20 to +70 CNon-operating maximum humidity 90% at 65 C (non-condensing)Non-operating maximum altitude 15,420 meters (50,000 feet)
31、07 | Keysight | E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors Data SheetRegulatory InformationElectromagnetic compatibility This product conforms with the protection requirements of European Council Directive 89/336/EEC for Electromagnetic Compatibility (E
32、MC). The conformity assessment requirements have been met using the technical Construction file route to compliance, using EMC test specifications EN 55011:1991 (Group 1, Class A) and EN 50082-1:1992. In order to preserve the EMC performance of the product, any cable which becomes worn or damaged mu
33、st be replaced with the same type and specification.Product safety This product conforms to the requirements of European Council Directive 73/23/EEC, and meets the following safety standards:IEC 61010-1(1990) + A1 (1992) + A2 (1995) / EN 61010-1 (1993)IEC 825-1 (1993) / EN 60825-1 (1994) Canada / CS
34、A C22.2 No. 1010.1-93Physical SpecificationsDimensions The following dimensions exclude front and rear panel protrusions: 212.6 mm W x 88.5 mmH x 348.3 mm D (8.5 in x 3.5 in x 13.7 in)WeightNet E4416A 4.0 kg (8.8 lbs) approximateE4417A 4.1 kg (9.0 lbs) approximateShipping E4416A 7.9 kg (17.4 lbs) ap
35、proximateE4417A 8.0 kg (17.6 lbs) approximateOrdering InformationStandard-shipped accessoriesPower sensor cableE9288A 1.5 meter (5 ft). One per E4416A, two per E4417APower cord One 2.4 meter (7.5 ft) cable. Power plug matches destination requirements.Manuals Product CD-ROM (contains English and loca
36、lized Users Guide and Programming Guide)Warranty Included with each EPM-P power meter is a standard 12-month return-to-Keysight warranty and service plan. A selection can be made to extend the initial warranty and service plan to 3 or 5 years. Standard-shipped accessories come with a 3-month warrant
37、y.08 | Keysight | E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors Data SheetPower Meter OptionsConnectorsE441xA-002 Parallel rear panel sensor input connector(s) and front panel reference calibrator connectorE441xA-003 Parallel rear panel sensor input connec
38、tor(s) and rear panel reference calibrator connectorCalibration documentationE441xA-A6J NSI/ NCSL Z540-1-1994 compliant calibration test data including measurement uncertaintiesE441x A-1A7 ISO/ IEC 17025:2005 compliant calibration test data including measure-ment uncertaintiesDocumentationE441xA-0BF
39、 Hard copy English language Programming GuideE441xA-0BK Hard copy English language Users Guide and Programming GuideE441xA-0B3 Hard copy English language Service ManualE441xA-ABD Hard copy German localization Users Guide and Programming GuideE441xA-ABE Hard copy Spanish localization Users Guide and
40、Programming GuideE441xA-ABF Hard copy French localization Users Guide and Programming GuideE441xA-ABJ Hard copy Japanese localization Users Guide and Programming GuideE441xA-ABZ Hard copy Italian localization Users Guide and Programming GuidePower sensor cablesE441xA-004 Delete power sensor cableFor
41、 operation with the E9320 power sensors:E9288A Power sensor cable, length 5 ft (1.5 m) E9288B Power sensor cable, length 10 ft (3 m) E9288C Power sensor cable, length 31 ft (10 m)Note: The E9288A, B, and C sensor cables will also operate with 8480, N8480 and E-series power sensors.For operation with
42、 8480, N8480, E-series CW and E9300 power sensors:11730A Power sensor and SNS noise source cable, length 5 ft (1.5 m)11730B Power sensor and SNS noise source cable, length 10 ft (3 m)11730C Power sensor and SNS noise source cable, length 20 ft (6.1 m)11730D Power sensor cable, length 50 ft (15.2 m)1
43、1730E Power sensor cable, length 100 ft (30.5 m)11730F Power sensor cable, length 200 ft (61.0 m)Other sensor cable lengths can be supplied on request.AccessoriesE441xA-908 Rack mount kit (one instrument)E441xA-909 Rack mount kit (two instruments)34131A Transit case for half-rack 2U high instruments
44、34141A Yellow soft carry/operating case34161A Accessory pouch09 | Keysight | E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors Data SheetService OptionsWarranty and calibration1R-50C-011-3 Keysight Calibration Upfront Plan 3-year coverageR-50C-011-5 Keysight C
45、alibration Upfront Plan 5-year coverageR-51B-001-3C 1 year Return-to-Keysight warranty extended to 3 yearsR-51B-001-5C 1 year Return-to-Keysight warranty extended to 5 yearsThe E9320 Series power sensors have a 12-month Return-to-Keysight warranty and service plan. For more information, contact your
46、 local sales and service office.E-Series E9320 Power Sensor SpecificationsThe E9320 peak and average power sensors are designed for use with the EPM-P series power meters. The E9320 sensors have two measurement modes:Normal mode Default mode for E9320 sensors) for peak, average and time-related meas
47、urementsAverage only mode Designed primarily for average power measurements on low-level signals. This mode is the only mode used with 8480 and N8480 series sensors, E-series CW sensors and E-series E9300 sensors.The following specifications are valid after zero and calibration of the power meter.No
48、te: E9320 power sensors MUST be used with an E9288A, B or C cable.1. Options not available in all countries.2. For average power measurements, free run acquisition.Table 3. Sensor specificationSensor model Video bandwidth Frequency range Power range Maximum power Connector typeAverage only mode Norm
49、al mode2E9321AE9325A300 kHz 50 MHz to 6 GHz50 MHz to 18 GHz65 dBm to +20 dBm 50 dBm to +20 dBm +23 dBm average; +30 dBm peak ( 10 sec duration)Type N (m)E9322AE9326A1.5 MHz 50 MHz to 6 GHz50 MHz to 18 GHz60 dBm to +20 dBm 45 dBm to +20 dBmE9323AE9327A5 MHz 50 MHz to 6 GHz50 MHz to 18 GHz60 dBm to +20 dBm 40 dBm to +20 dBm10 | Keysight | E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors Data SheetTable 4. Lower and upper measurement rangesE9321A